System, X-Ray, Tomography, Computed. 510(k) K081105, from GE Healthcare (Waukesha, WI). Product code JAK, class 2.
As of Nov 17, 2026, K081105 is an FDA 510(k) held by GE Healthcare, substantially equivalent on May 9, 2008 after 21 days, with 5 recalls tied to it in FDA's recall database.
- Substantially equivalent: May 9, 2008
- Review time: 21 days
- Recalls: 5
- Cited as a predicate by: 5 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe GE LightSpeed CT750 HD Computed Tomography X-ray system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, Spectral, and Gated acquisitions for all ages. These images may be obtained either with or without contrast.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Lightspeed 7.1 Ct Scanner SystemK061817 · Ge Medical Systems, LLC · Jul 12, 2006
Cited as a Predicate
5 later 510(k) summaries name this device as a predicate- Revolution CTK163213 · Ge Medical Systems, LLC · Dec 16, 2016
- Asir-VK133640 · Ge Medical Systems, LLC · Mar 25, 2014
- Discovery CT750 HdK120833 · Ge Medical Systems, LLC · Jun 12, 2012
- Ge Veo Reconstruction OptionK103489 · General Electric Co. · Sep 9, 2011
- Brilliance Dual Energy OptionK090462 · Philips Medical Systems (Cleveland), Inc. · Jun 23, 2009
Recalls
Recalls FDA ties to K081105, newest first; each opens FDA's recall record| Nov 5, 2024 | Z-0616-2025GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical,… | Open, Classified |
| Nov 5, 2024 | Z-0615-2025GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical,… | Open, Classified |
| Jul 25, 2011 | Z-1188-2013Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil… | Terminated |
| Apr 5, 2011 | Z-1940-2013GE Healthcare conducted a recall involving 2 events affecting the GE LightSpeed CT750 HD. Event 1 - There was a report regarding CT number variance in time line at posterior fossa region of head perfusion. When beam… | Terminated |
| Mar 25, 2011 | Z-1148-2014When performing a secondary capture generated from GSI Viewer, the header information is correct on the images. However, when changing the ROI (region of interest) and zooming in, the DICOM header does not change to… | Terminated |
Other JAK Clearances
The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Apr 18, 2008
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files