Arca
Skip to content

Fresenius Model 2008T Hemodialysis Machine

Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K080964, from Fresenius Medical Care North America (Waltham, MA). Product code KDI, class 2.

As of Nov 17, 2026, K080964 is an FDA 510(k) held by Fresenius Medical Care North America, substantially equivalent on Jan 14, 2009 after 285 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jan 14, 2009
  • Review time: 285 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Fresenius 2008T represents the next genesis of the Fresenius 2008K model hemodialysis machine. Hemodialysis is prescribed by physicians for patients with acute or chronic renal failure, when conservative therapy is judged inadequate. Dialysis therapy may be intermittent or continuous.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K080964, newest first; each opens FDA's recall record
Oct 5, 2017Z-0829-2018While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate.…Terminated
Mar 10, 2016Z-1325-2016Added Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)Terminated
Feb 20, 2015Z-1550-2015The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.Terminated
May 24, 2013Z-1693-2013Incorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)Terminated

Other KDI Clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Special
Received
Apr 4, 2008
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.