Surgical Handpieces, Mikro Saw Handpieces
Handpiece, Rotary Bone Cutting. 510(k) K080939, from W&H Dentalwerk Buermoos GmbH (Buermoos, AT). Product code KMW, class 2.
As of Nov 17, 2026, K080939 is an FDA 510(k) held by W&H Dentalwerk Buermoos GmbH, substantially equivalent on Jul 29, 2008 after 117 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Jul 29, 2008
- Review time: 117 days
- Recalls: 1
- Cited as a predicate by: 2 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Surgical Contra-Angel Handpieces Types WS-56E,WS-75E/KM,WS-92 E/3,SURGICAL Straight Handpieces Types, S-9, S-10, S-11K011061 · W&H Dentalwerk Buermoos GmbH · Jun 28, 2001
- Implantmed SI-915 (115V Version); Implantmed SI-923 (230V Version) Incl. AccessoriesK052741 · W&H Dentalwerk Buermoos GmbH · Jan 13, 2006
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- W&H Implantmed SI-1015 incl. AccessoriesK161957 · W&H Dentalwerk Burmoos GmbH · Nov 23, 2016
- Montblanc Surgical Contra-Angle And Straight HandpieceK111532 · Anthogyr Sas · Oct 7, 2011
Recalls
Recalls FDA ties to K080939, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Jan 16, 2024 | Z-1168-20241:2 speed increasing surgical handpieces have been incorrectly laser marked during production. The specific laser marking should read 1:2, but instead it reads 1:1 on the body of the handpiece | Open, Classified |
Other KMW Clearances
The newest 510(k)s for the same product code; see every KMW device- Oral Surgery ContraK211584 · Nakanishi, Inc. · Aug 19, 2021
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 3, 2008
- Review panel
- Dental
- Regulation
- 21 CFR 872.4120
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files