Modeification To Iplan RT Dose

System, Planning, Radiation Therapy Treatment. 510(k) K080888, from Brainlab AG (Feldkirchen, DE). Product code MUJ, class 2.

As of Nov 20, 2026, K080888 is an FDA 510(k) held by Brainlab AG, substantially equivalent on Jun 11, 2008 after 72 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 72 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

iPlan RT Dose is a stereotactic radiation treatment planning system that is intended for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • Iplan RTK103246 · Brainlab AG · Mar 16, 2011

Recalls

Recalls FDA ties to K080888, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 19, 2014Z-0957-2015, Z-0956-2015iPlan RT Radiation Treatment Planning Software: Potentially incorrect patient positioning when using multiple localized CT image data sets.Terminated
Aug 12, 2009Z-2251-2009Failure to conduct important safety checks when using the BrainLab radiotherapy treatment planning software in combination with BrainLab conical collimators could result in unintended radiation outside the conical…Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Special
Received
Mar 31, 2008
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files