Apexpro Telemetry Systems
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K080251, from Ge Medical Systems Information Technologies (Milwaukee, WI). Product code MHX, class 2.
As of Nov 18, 2026, K080251 is an FDA 510(k) held by Ge Medical Systems Information Technologies, substantially equivalent on Jun 20, 2008 after 141 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 141 days
- Recalls: 4
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe ApexPro Telemetry System is intended for use under the direct supervision of a licensed healthcare practitioner. The system is designed to acquire and monitor physiological data for ambulating patients within a defined coverage area. The system processes this physiological data to detect various ECG arrhythmia events and select physiological parameter limit violations.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Modification To: Apexpro Telemetry SystemK032369 · Ge Medical Systems Information Technologies · Aug 15, 2003
- Apexpro FH Telemetry SystemK033365 · Ge Medical Systems Information Technologies · Nov 6, 2003
- Apex OximeterK980299 · Marquette Medical Systems, Inc. · Apr 24, 1998
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Surveyor S4 Mobile MonitorK160685 · Mortara Instrument, Inc. · Aug 4, 2016
Recalls
Recalls FDA ties to K080251, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Apr 24, 2026 | Z-2242-2026GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF… | Open |
| Nov 12, 2019 | Z-0725-2020When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and… | Terminated |
| Aug 30, 2019 | Z-0175-2020May not provide visual and/or audible alarms at the CARESCAPE Central Station or Clinical Information Center monitor for ECG arrhythmias, ECG LEADS FAIL or Pulse Oximetry (SpO2) under certain conditions. | Open |
| Jan 28, 2011 | Z-1522-2011GE Healthcare has become aware of a potential for loss of patient monitoring when ApexPro Telemetry System is used in conjunction with a mirrored view CARESCAPE CIC Pro in combination monitoring mode (combo mode). The… | Terminated |
Other MHX clearances
The newest 510(k)s for the same product code; their median review was 207 days; see every MHX device- 12SL PrecisionK261307 · Ge Medical Systems Information Technologies, Inc. · Nov 3, 2026
- Radius VSM and AccessoriesK260676 · Masimo Corporation · Oct 14, 2026
- Carescape ONE; Carescape Dock F0; TruSignal ECG MM; TruSignal PRES MM; TruSignal TEMP MM; TruSignal SpO2 MM 1.8M;…K260656 · GE Healthcare Finland Oy · Sep 1, 2026
- Carescape Canvas 1000; Carescape Canvas Smart DisplayK260343 · GE Healthcare Finland Oy · Aug 28, 2026
- Infinity® Acute Care System (IACS) – VG8.1K260579 · Draeger Medical Systems, Inc. · Aug 20, 2026
- Telemetry Monitor (iT30, iT50)K251508 · Edan Instruments, Inc. · Aug 6, 2026
- Monitor B105M; Monitor B125M; Monitor B155M; Monitor B105P; Monitor B125PK253277 · Ge Medical Systems Information Technologies, Inc. · Jul 24, 2026
- Radius VSM and AccessoriesK254209 · Masimo Corporation · Jun 4, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 31, 2008
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files