System, X-Ray, Tomography, Computed. 510(k) K080211, from Toshiba America Medical Systems, In.C (Tustin, CA). Product code JAK, class 2.
As of Nov 19, 2026, K080211 is an FDA 510(k) held by Toshiba America Medical Systems, In.C, substantially equivalent on Mar 14, 2008 after 45 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 45 days
- Recalls: 8
- Cited as a predicate by: 2 devices
- TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT ScannerK051833 · Toshiba America Medical Systems, In.C · Jul 15, 2005
- Aquilion CxlK112989 · Toshiba America Medical Systems, In.C · Apr 10, 2012
- Scenaria Whole-Body X-Ray CT SystemK101888 · Hitachi Medical Systems America, Inc. · Apr 8, 2011
| May 8, 2015 | Z-0013-2016If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved. | Terminated |
| Oct 1, 2013 | Z-1113-2014Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the… | Terminated |
| Dec 8, 2010 | Z-3221-2011There is the potential that the current may not be controlled correctly in examinations using the ECG modulation function, which adjusts the x-ray output in synchronization with the ECG waveform. | Terminated |
| Apr 9, 2010 | Z-2204-2011The recall was initiated by Toshiba America Medical Systems (TAMS) because the software associated with TSX-101A: Aquilion 64 can prevent scanning from being performed at the specified helical pitch (couch movement… | Terminated |
| Jan 8, 2009 | Z-1168-2010, Z-1166-2010, Z-1167-2010, Z-1169-2010Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10… | Terminated |
- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Jan 29, 2008
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files