Model TSX-101A/H/I Aquilion 32/64 SP CT System

System, X-Ray, Tomography, Computed. 510(k) K080211, from Toshiba America Medical Systems, In.C (Tustin, CA). Product code JAK, class 2.

As of Nov 19, 2026, K080211 is an FDA 510(k) held by Toshiba America Medical Systems, In.C, substantially equivalent on Mar 14, 2008 after 45 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 45 days
  • Recalls: 8
  • Cited as a predicate by: 2 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K080211, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 8, 2015Z-0013-2016If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved.Terminated
Oct 1, 2013Z-1113-2014Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the…Terminated
Dec 8, 2010Z-3221-2011There is the potential that the current may not be controlled correctly in examinations using the ECG modulation function, which adjusts the x-ray output in synchronization with the ECG waveform.Terminated
Apr 9, 2010Z-2204-2011The recall was initiated by Toshiba America Medical Systems (TAMS) because the software associated with TSX-101A: Aquilion 64 can prevent scanning from being performed at the specified helical pitch (couch movement…Terminated
Jan 8, 2009Z-1168-2010, Z-1166-2010, Z-1167-2010, Z-1169-2010Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10…Terminated

Other JAK clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Traditional, third-party review
Received
Jan 29, 2008
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files