GE Datex-Ohmeda Aisys
Gas-Machine, Anesthesia. 510(k) K073707, from Datex-Ohmeda, Inc. (Madison, WI). Product code BSZ, class 2.
As of Nov 19, 2026, K073707 is an FDA 510(k) held by Datex-Ohmeda, Inc., substantially equivalent on Jan 30, 2008 after 30 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 30 days
- Recalls: 2
- Cited as a predicate by: 3 devices
- GE Datex-Ohmeda S/5 Adu CarestationK090892 · GE Healthcare Finland Oy · Apr 30, 2009
- GE Datex-Ohmeda AisysK090233 · Datex-Ohmeda, Inc. · Mar 5, 2009
- GE Datex-Ohmeda AvanceK081844 · Datex-Ohmeda, Inc. · Aug 29, 2008
| Feb 5, 2009 | Z-1425-2009Unintended shut-down: The anesthesia machine may start shutting down without human intervention due to a faulty On/Standby Switch. Prior to shutdown, the system will issue an alarm and play a message that shutdown will… | Terminated |
| Aug 28, 2008 | Z-0020-2009The electronic vaporization system in the GE Aisys Anesthesia machine contains the following components: Backpressure Valve, Inflow Check Valve and Cassette Interface Board, that can contribute to independent failures. | Terminated |
- A5 Anesthesia System (A5); A7 Anesthesia System (A7)K253330 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Jun 25, 2026
- Atlan (A100); Atlan (A100 XL); Atlan(A350); Atlan (A350 XL)K252746 · Drägerwerk AG Co. Kgaa · May 21, 2026
- Codonics Safe Labeling System-Point of Care Station (PCS) (SLS-700i and related 610i, 620i, 630i series enabled to…K251352 · Codonics.Incorporated · Jul 11, 2025
- AtlanK230931 · Drägerwerk AG Co. Kgaa · Jul 23, 2023
- Carestation 750/750cK213867 · Datex-Ohmeda, Inc. · Mar 28, 2023
- A8, A9 Anesthesia SystemK201957 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Mar 26, 2021
- Type
- Special
- Received
- Dec 31, 2007
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5160
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files