Arca
Skip to content
Tools
Tools

Remstar Pro M-Series Cpap & Heated Humidifier System

Ventilator, Non-Continuous (Respirator). 510(k) K072996, from Respironics, Inc. (Murrysville, PA). Product code BZD, class 2.

As of Nov 21, 2026, K072996 is an FDA 510(k) held by Respironics, Inc., substantially equivalent on Dec 27, 2007 after 64 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 64 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Intended Use/Indications for Use The Respironics REMstar Pro M-Series CPAP & Heated Humidifier System delivers positive airway pressure therapy for the treatment of Obstructive Sleep Apnea in spontaneously breathing patients weighing over 30kg. Environment of Use/Patient Population For use in the home or hospital/institutional environment.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K072996, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Feb 5, 2009Z-1260-2009Malfunction of the J3 connection between the Heater Plate Main printed circuit assembly (PCA) and the heater plate may have an intermittent connection which may lead to excessive heat build-up on the humidifier PCA.Terminated

Other BZD clearances

The newest 510(k)s for the same product code; their median review was 263 days; see every BZD device

Record

Type
Special
Received
Oct 24, 2007
Review panel
Anesthesiology
Regulation
21 CFR 868.5905

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files