Arca
Skip to content
Tools
Tools

Acuson X300 Ultrasound System

System, Imaging, Pulsed Echo, Ultrasonic. 510(k) K072676, from Siemens Medical Solutions USA, Inc. (Mountain View, CA). Product code IYO, class 2.

As of Nov 19, 2026, K072676 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Dec 18, 2007 after 88 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 88 days
  • Recalls: 5
  • Cited as a predicate by: 3 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K072676, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 17, 2010Z-1838-2011There is an error in the estimation of mean pressure gradient in the Cardiac measurement package. The currently implemented calculation will result in an underestimation of the mean pressure gradient (PGmean) when…Terminated
Jul 30, 2009Z-2065-2011Software Error with potential for misinterpretation of data-- The previous patient's last capture clip image can appear in the next patient's study.Terminated
Jul 24, 2009Z-2061-2011Product Failure-- When the 64-Channel System V5M is used in the 128-Channel mode, an over current fault occurs which shuts the system down.Terminated
Jul 24, 2009Z-2147-2011Mistreatment risk-- when a user defined formula is used to calculate a value in a vascular exam, data is switched laterally when it is internally transferred and displayed on the summary report within the system.Terminated
Aug 19, 2008Z-0111-2009Thermal Index cranial (TIC) is not displayed for the Neo-Head exam type with the C8-5 transducer. This is a required display to support the alara principle when imaging cranial structure.Terminated

Other IYO clearances

The newest 510(k)s for the same product code; their median review was 234 days; see every IYO device

Record

Type
Special
Received
Sep 21, 2007
Review panel
Radiology
Regulation
21 CFR 892.1560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files