Arca
Skip to content
Tools
Tools

Sequoia Spinal System, Model 3300 Series

Thoracolumbosacral Pedicle Screw System. 510(k) K072672, from Abbott Spine, Inc. (Austin, TX). Product code NKB, class 2.

As of Nov 18, 2026, K072672 is an FDA 510(k) held by Abbott Spine, Inc., substantially equivalent on Nov 28, 2007 after 68 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 68 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

When intended for pedicle screw fixation from T1 –S1, the Sequoia Thoracolumbar Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (defined as…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K072672, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 4, 2011Z-2705-2011A rounded or partially rounded tip on the screwdriver may not provide the required locking torque to lock the closure top to the polyaxial screw. When the hex tip of the screwdriver is not fully engaged and/or the…Terminated
Feb 13, 2008Z-1479-2008Screw Head Dissociation: The polyaxial screw head may dissociate from screw shaft during surgery. Also, the polyaxial screw driver is experiencing difficulty (broken/bent driver shafts, difficulty attaching screws to…Terminated

Other NKB clearances

The newest 510(k)s for the same product code; their median review was 90 days; see every NKB device

Record

Type
Special
Received
Sep 21, 2007
Review panel
Orthopedic
Regulation
21 CFR 888.3070

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files