Aquilion One TSX-301A/2 Computed Tomography System
System, X-Ray, Tomography, Computed. 510(k) K072606, from Toshiba America Medical Systems, In.C (Tustin, CA). Product code JAK, class 2.
As of Nov 19, 2026, K072606 is an FDA 510(k) held by Toshiba America Medical Systems, In.C, substantially equivalent on Oct 2, 2007 after 15 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 15 days
- Recalls: 4
- Cited as a predicate by: 1 device
This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc. The Aquilion ONE has the capability to provide volume sets of the entire organ.
- Model TSX-301B Aquilion CT SystemK083282 · Toshiba America Medical Systems, In.C · Dec 12, 2008
| Jun 22, 2011 | Z-3196-2011Toshiba America Medical Systems is initiating a field correction on their TSX-301B: AQUILION Premium; System, X-Ray, Tomography, Computed, TSX-301A: AQUILION One; System, X-Ray, Tomography, Computed, TSX-20IA: AQUlLlON… | Terminated |
| Jul 13, 2009 | Z-2282-20121. It has been found that in raw data processing for Go&Return helical scanning, when the raw data acquired in the first scan (Go) and that acquired in the second scan (Return) are selected and reconstructed at one… | Terminated |
| Feb 3, 2009 | Z-2339-2010ECG Gated Dynamic Volume Prospective scan may stop if preset eXam Plan is confirmed and CTA On is changed to CTA OFF, due to exposure control PWB. | Terminated |
| May 13, 2008 | Z-2468-2012The SureStart function may not operate in the usual manner, and you may obtain abnormal analysis results when using the cerebral blood flow analysis function. | Terminated |
- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Sep 17, 2007
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files