Integra Radionics Interfix Patient Adapters
Accelerator, Linear, Medical. 510(k) K072454, from Integra Radionics, Inc. (Burlington, MA). Product code IYE, class 2.
As of Nov 20, 2026, K072454 is an FDA 510(k) held by Integra Radionics, Inc., substantially equivalent on Oct 16, 2007 after 46 days, with 1 recall tied to it in FDA's recall database.
- Review time: 46 days
- Recalls: 1
For cranial fixation during CT scanning and treatment with the TomoTherapy HiArt System.
- LCM-2-Linac Couch-Mount SystemK924632 · Radionics, Inc. · Apr 7, 1993
- Xknife Radiosurgery SystemK912630 · Rsa, Inc. · Nov 4, 1991
| Started | Recall | Status |
|---|---|---|
| Jan 23, 2012 | Z-1187-2012Material used in the superior region of the InterFix CT Adapter (CTA) has a different density than that of the Treatment Table Couch Adapter (TTA). Resulting dosimetry can be affected if posterior fields are used.… | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 31, 2007
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files