Arca
Skip to content
Tools
Tools

Durex Synthetic Polyisoprene Male Condom

Condom, Synthetic. 510(k) K072169, from Ssl Americas, Inc. (Norcross, GA). Product code MOL, class 2.

As of Nov 26, 2026, K072169 is an FDA 510(k) held by Ssl Americas, Inc., substantially equivalent on Jun 19, 2008 after 318 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 318 days
  • Recalls: 4

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Durex synthetic polyisoprene male condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted disease).

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K072169, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 19, 2018Z-0993-2019, Z-0994-2019Five batches of the product have failed the specification of 1.0kPa for Burst Pressure.Terminated
Aug 6, 2018Z-0091-2019Product marketed in U.S. does not have sufficient data to ensure that it will meet the internal burst pressure specifications at the end of 5 yrs shelf-lifeTerminated
Jul 23, 2018Z-3241-2018The products failed the filed specification of 1.0kPa for Burst Pressure.Terminated

Other MOL clearances

The newest 510(k)s for the same product code; their median review was 264 days; see every MOL device

Record

Type
Abbreviated
Received
Aug 6, 2007
Review panel
Obstetrics/Gynecology
Regulation
21 CFR 884.5300

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.