P.004 Rc Cares Titanium And Ceramic Abutments
Abutment, Implant, Dental, Endosseous. 510(k) K072151, from Straumann USA (Andover, MA). Product code NHA, class 2.
As of Nov 17, 2026, K072151 is an FDA 510(k) held by Straumann USA, substantially equivalent on Sep 11, 2007 after 39 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Sep 11, 2007
- Review time: 39 days
- Recalls: 1
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFAbutments are placed into the dental implants to provide support for prosthetic restoration such as crowns or bridges. The P.004 RC CARES Titanium Abutment is indicated for cemented restoration. The abutment can be used in single tooth replacements and multiple tooth restorations.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Straumann Computer Aided Restoration ServiceK052272 · Straumann USA · Jan 3, 2006
- Straumann Computer Aided Restoration Service (C.A.R.E.S.) Ceramic CopingK061277 · Straumann USA · Apr 12, 2007
- P.0004 ImplantsK062129 · Institut Straumann AG · Oct 20, 2006
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- TruAbutment DSK172304 · Truabutment, Inc. · Jan 23, 2018
- Straumann CARES Titanium Alloy (TAN) AbutmentK150899 · Straumann USA, LLC · Sep 16, 2015
- Zimmer Zfx Titanium Abutment For Straumann Bone Level Implant SystemK141120 · Zimmer Dental, Inc. · Jul 29, 2014
- P.004 Nc Cares Titanium And Ceramic AbutmentsK081005 · Institut Straumann AG · May 2, 2008
Recalls
Recalls FDA ties to K072151, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Mar 17, 2026 | Z-1632-2026A material mix-up had occurred during the manufacturing of Straumann Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti… | Open, Classified |
Other NHA Clearances
The newest 510(k)s for the same product code; their median review was 169 days; see every NHA device- Elos Accurate Hybrid BaseK261327 · Elos Medtech Pinol A/S · Sep 30, 2026
- Cemented Plus AbutmentK262207 · Unidental Co., Ltd. · Sep 23, 2026
- BellaTek BarsK260940 · ZimVie US Corp, LLC · Sep 21, 2026
- Universal Base ASC Engaging; Universal Base ASC Non-Engaging; Universal Base MUA ASC;…K253999 · Nobel Biocare AB · Aug 6, 2026
- Neodent Implant System - NeoConvert FitK261147 · JJGC Indústria e Comércio de Materiais Dentários S.A. · Jul 1, 2026
- Mist IcK261122 · Imagine Milling Technologies, LLC · Jun 27, 2026
- BioHorizons CEREC Compatible Ti-BasesK260061 · BioHorizons Implant Systems, Inc. · Jun 26, 2026
- Neodent Scannable Healing AbutmentK253922 · JJGC Indústria e Comércio de Materiais Dentários S.A. · Jun 15, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 3, 2007
- Review panel
- Dental
- Regulation
- 21 CFR 872.3630
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files