Arca
Skip to content
Tools
Tools

Normoflo Irrigation Fluid Warmer, Models H-1100, H-1129, Normoflo Irrigation Warming Set, Models, IR-40, IR-500, IR-600,

Warmer, Thermal, Infusion Fluid. 510(k) K072080, from Smiths Medical Asd, Inc. (Rockland, MA). Product code LGZ, class 2.

As of Nov 19, 2026, K072080 is an FDA 510(k) held by Smiths Medical Asd, Inc., substantially equivalent on Jun 2, 2008 after 308 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 308 days
  • Recalls: 2
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Level 1® NORMOFLO® Irrigation Warming Set (Disposable Set) is a disposable irrigating fluid administration set, designed for use by trained medical professionals, for warming irrigating fluids. The Disposable Set is intended for use with the Level 1® H-1100 series NORMOFLO® Irrigation Fluid Warmers only.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K072080, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Aug 6, 2021Z-2377-2021, Z-2376-2021There is potential for harmful levels of aluminum leaching when using lower flow rates with certain solutions and blood products and longer duration of use.Open

Other LGZ clearances

The newest 510(k)s for the same product code; their median review was 322 days; see every LGZ device

Record

Type
Traditional
Received
Jul 30, 2007
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files