Arca
Skip to content

Metha Hip System

Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate. 510(k) K071916, from Aesculap Implant Systems, Inc. (Center Valley, PA). Product code MEH, class 2.

As of Nov 17, 2026, K071916 is an FDA 510(k) held by Aesculap Implant Systems, Inc., substantially equivalent on Jun 2, 2008 after 327 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Jun 2, 2008
  • Review time: 327 days
  • Recalls: 1
  • Cited as a predicate by: 7 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Metha® Hip System (uncemented, press-fit fixation) is intended to replace a hip joint. The device is intended for: - skeletally mature individuals undergoing primary surgery for total hip replacement - patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K071916, newest first; each opens FDA's recall record
StartedReasonStatus
Jul 19, 2011Z-3014-2011Aesculap Implant Systems received a notification from the manufacturer, Aesculap, AG of a fractured CoCr neck adapter.Terminated

Other MEH Clearances

The newest 510(k)s for the same product code; their median review was 90 days; see every MEH device

Record

Type
Traditional
Received
Jul 11, 2007
Review panel
Orthopedic
Regulation
21 CFR 888.3353

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.