Zodiac 4.0 Polyaxial Spinal Fixation System
Orthosis, Spondylolisthesis Spinal Fixation. 510(k) K071890, from Alphatec Spine, Inc. (Carlsbad, CA). Product code MNH, class 2.
As of Nov 17, 2026, K071890 is an FDA 510(k) held by Alphatec Spine, Inc., substantially equivalent on Nov 9, 2007 after 123 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Nov 9, 2007
- Review time: 123 days
- Recalls: 1
- Cited as a predicate by: 5 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFIt is intended that this device, in any system configuration be removed after development of solid fusion mass. Hook component indications are limited to T7-L5. Sacral screw indications are limited to the sacrum only.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Modification To Moss Miami Spinal System Polyaxial ScrewsK030383 · Depuyacromed · Feb 26, 2003
Cited as a Predicate
5 later 510(k) summaries name this device as a predicate- Kodiak Spinal Fixation SystemK181677 · Alphatec Spine, Inc. · Aug 9, 2018
- DPZ Pedicular Fixation SystemK150294 · Osteomed Implantes, Ltda · Jun 5, 2015
- Cosmolock Pedicle Screw SystemK140678 · Kalitec Direct, LLC · Jun 19, 2014
- Illico Mis Posterior Fixation SystemK123623 · Alphatec Spine, Inc. · Feb 14, 2013
- Xenon(Tm) Pedicle Screw SystemK111634 · Alphatec Spine, Inc. · Jan 25, 2012
Recalls
Recalls FDA ties to K071890, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Sep 25, 2012 | Z-0168-2013The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated Polyaxial Screws, therefore rendering… | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 9, 2007
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3070
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files