Aurora Led Series Surgical Lights
Light, Surgical, Ceiling Mounted. 510(k) K071698, from Skytron, Div. the Kmw Group, Inc. (Grand Rapids, MI). Product code FSY, class 2.
As of Nov 28, 2026, K071698 is an FDA 510(k) held by Skytron, Div. the Kmw Group, Inc., substantially equivalent on Jul 6, 2007 after 15 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 15 days
- Recalls: 4
- Cited as a predicate by: 2 devices
The SKYTRON Aurora series surgical light fixture is intended to be used by medical personnel for the purpose of illuminating surgical sites
- StellarK002463 · Skytron, Div. the Kmw Group, Inc. · Nov 8, 2000
- 21ST Century Centurion Exceled and System Two Led Lighting SystemsK100388 · Medical Illumination International, Inc. · Jun 22, 2010
- Nuvo Verde Led Lighting SystemK083323 · Nuvo, Inc. · Jan 28, 2009
| Started | Recall | Status |
|---|---|---|
| Jul 9, 2024 | Z-2620-2024, Z-2621-2024, Z-2619-2024, Z-2618-2024The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching. | Open |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Jun 21, 2007
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4580
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files