Arca
Skip to content
Tools
Tools

Aurora Led Series Surgical Lights

Light, Surgical, Ceiling Mounted. 510(k) K071698, from Skytron, Div. the Kmw Group, Inc. (Grand Rapids, MI). Product code FSY, class 2.

As of Nov 28, 2026, K071698 is an FDA 510(k) held by Skytron, Div. the Kmw Group, Inc., substantially equivalent on Jul 6, 2007 after 15 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 15 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The SKYTRON Aurora series surgical light fixture is intended to be used by medical personnel for the purpose of illuminating surgical sites

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • StellarK002463 · Skytron, Div. the Kmw Group, Inc. · Nov 8, 2000

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K071698, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jul 9, 2024Z-2620-2024, Z-2621-2024, Z-2619-2024, Z-2618-2024The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching.Open

Record

Type
Traditional, third-party review
Received
Jun 21, 2007
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.4580

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files