Stiffer Coaxial Micro-Introducer Kit
Introducer, Catheter. 510(k) K071574, from Enpath Medical, Inc. (Plymouth, MN). Product code DYB, class 2.
As of Nov 22, 2026, K071574 is an FDA 510(k) held by Enpath Medical, Inc., substantially equivalent on Jul 5, 2007 after 27 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 27 days
- Recalls: 2
- Cited as a predicate by: 2 devices
The Stiffer Coaxial Micro-Introducer Kit is indicated for use to introduce up to a 0.038 inch guide wire or catheter into the vascular systems following a small 21 gauge needle stick. Summary of Non-Clinical Testing Performance Testing: The performance testing for this device included testing to verify that the device functions in a safe and effective manner.
- Medamicus Coaxial Introducer Kit, 10330-XXX, Boston Scientific Coaxial Dilator SetK990705 · Medamicus, Inc. · Jun 28, 1999
- Vsi Micro-Introducer Sets (Regular, Stiffen), Models 7197V-7209V, 7240V-7256V, 7259V, 7230V, 7231V, 7260V, 7274VK101604 · Vascular Solutions, Inc. · Jan 24, 2011
- Micro-Stick SetK091954 · Medcomp · Jan 8, 2010
| Sep 8, 2016 | Z-0623-2017Greatbatch Medical has identified an issue with the Stiffer Coaxial Micro-Introducer 4FR and 5FR after a field complaint reported that the rotating luer detached from the dilator hub during use. | Terminated |
| Sep 8, 2016 | Z-0051-2017The supplier initiated a recall because affected lots of the micro-introducer have an outer diameter of the dilator hub that is slightly smaller than is capable of securing the dilator to the sheath | Terminated |
- Giranel Guiding Sheath SetK262674 · Kossel Medtech (Suzhou) Co., Ltd. · Oct 15, 2026
- NAMIC Dilator and SheathK253526 · Medline Industries, LP · Jul 29, 2026
- Micropuncture Kit / Micropuncture Access Kit / Micropuncture Introducer Set/ Microintroducer Kit / Microintroducer Kit…K261529 · Hartman Medical Group, LLC · Jul 6, 2026
- SimplGuideK261242 · Pendracare · Jun 25, 2026
- Health Line Tearaway IntroducerK261950 · Health Line International Corporation · Jun 22, 2026
- FieldFlex Steerable SheathK261492 · CenterPoint Systems, LLC · Jun 5, 2026
- Echo Large Bore Introducer SheathK260606 · Echo Medical, LLC · Apr 29, 2026
- Introducer Sheath SetK254248 · Cardiocycle Medical(Suzhou) Co., Ltd. · Apr 23, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jun 8, 2007
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1340
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files