Introducer, Catheter. 510(k) K071574, from Enpath Medical, Inc. (Plymouth, MN). Product code DYB, class 2.

As of Nov 22, 2026, K071574 is an FDA 510(k) held by Enpath Medical, Inc., substantially equivalent on Jul 5, 2007 after 27 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 27 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Stiffer Coaxial Micro-Introducer Kit is indicated for use to introduce up to a 0.038 inch guide wire or catheter into the vascular systems following a small 21 gauge needle stick. Summary of Non-Clinical Testing Performance Testing: The performance testing for this device included testing to verify that the device functions in a safe and effective manner.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K071574, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 8, 2016Z-0623-2017Greatbatch Medical has identified an issue with the Stiffer Coaxial Micro-Introducer 4FR and 5FR after a field complaint reported that the rotating luer detached from the dilator hub during use.Terminated
Sep 8, 2016Z-0051-2017The supplier initiated a recall because affected lots of the micro-introducer have an outer diameter of the dilator hub that is slightly smaller than is capable of securing the dilator to the sheathTerminated

Other DYB clearances

The newest 510(k)s for the same product code; their median review was 71 days; see every DYB device

Record

Type
Special
Received
Jun 8, 2007
Review panel
Cardiovascular
Regulation
21 CFR 870.1340

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.