Arca
Skip to content
Tools
Tools

Subtalar Implant

Screw, Fixation, Bone. 510(k) K071498, from Biomet Sports Medicine, Inc. (Warsaw, IN). Product code HWC, class 2.

As of Nov 20, 2026, K071498 is an FDA 510(k) held by Biomet Sports Medicine, Inc., substantially equivalent on Jul 31, 2007 after 61 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 61 days
  • Recalls: 7
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Subtalar Implant is indicated for use in treating the hyperpronated foot by stabilizing the subtalar joint. It is intended to block forward, downward, and medial displacement of the talus, thereby limiting the excessive eversion of the hindfoot.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K071498, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jan 21, 2011Z-1614-2011, Z-1618-2011, Z-1613-2011, Z-1616-2011, Z-1617-2011, Z-1615-2011, Z-1612-2011The firm has initiated this recall following internal fatigue testing that shows possible initiation of fatigue cracking in implants under load.Terminated

Other HWC clearances

The newest 510(k)s for the same product code; their median review was 157 days; see every HWC device

Record

Type
Traditional
Received
May 31, 2007
Review panel
Orthopedic
Regulation
21 CFR 888.3040

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files