Arca
Skip to content
Tools
Tools

Varisource Ix Hdr Afterloader

System, Applicator, Radionuclide, Remote-Controlled. 510(k) K071467, from Varian Medical Systems, Inc. (Palo Alto, CA). Product code JAQ, class 2.

As of Nov 20, 2026, K071467 is an FDA 510(k) held by Varian Medical Systems, Inc., substantially equivalent on Jul 16, 2007 after 48 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 48 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The VariSource iX is indicated, in the treatment of both benign and malignant disease, for both curative and palliative intent, in the delivery of remote-controlled high dose-rate Brachytherapy for conditions anywhere in the body when brachytherapy treatment is indicated.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K071467, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 5, 2012Z-1199-2012, Z-1200-2012Components of medical device used in brachytherapy may be defective and cause a shifted dose distribution to patients.Terminated
Mar 1, 2010Z-0325-2012VariSource iX console software may cause incorrect dwell times to be delivered to patient and may result in a mistreatment.Terminated

Other JAQ clearances

The newest 510(k)s for the same product code; their median review was 98 days; see every JAQ device

Record

Type
Traditional
Received
May 29, 2007
Review panel
Radiology
Regulation
21 CFR 892.5700

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files