Optical Guidance Platform
Accelerator, Linear, Medical. 510(k) K071360, from Varian Medical Systems, Inc. (Palo Alto, CA). Product code IYE, class 2.
As of Nov 25, 2026, K071360 is an FDA 510(k) held by Varian Medical Systems, Inc., substantially equivalent on Jun 8, 2007 after 24 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 24 days
- Recalls: 5
Varian Medical System’s Optical Guidance Platform (OGP) is for use with a charged particle accelerator to perform precise positioning of treatment target for stereotactic radiosurgery or radiotherapy treatments on cranial or extracranial lesions.
- Snt Stereotactic Localization SystemK971675 · Surgical Navigation Technologies, Inc. · Jul 14, 1997
- Snt Linac AccessoriesK971893 · Surgical Navigation Technologies, Inc. · Nov 13, 1997
- Radiocameras Patient Positioning SystemK980750 · Surgical Navigation Technologies, Inc. · May 20, 1998
- Radiocameras Biteblock Localization and Positioning SystemK981346 · Surgical Navigation Technologies, Inc. · Jul 10, 1998
- Radiocameras System Head/Neck ApplicationK994355 · Surgical Navigation Technologies, Inc. · Mar 22, 2000
- Radiocameras Extracranial SystemK000246 · Surgical Navigation Technologies, Inc. · Oct 20, 2000
| Oct 4, 2011 | Z-0332-2012An anomaly has been identified with the Head Ring posts used by both the Optical Guidance Platform FrameArray module and the Floorstand where the head ring posts may be damaged due to excessive mechanical stress… | Terminated |
| Mar 30, 2011 | Z-2200-2011A software anomaly has been identified with the Optical Guidance Platform (OGP) Software v2.6 and v2.6.1 where the transfer of datasets from treatment planning systems other than the FastPlan system result in a lateral… | Terminated |
| Mar 3, 2011 | Z-2063-2011The optical guidance platform may not be properly enforcing a 24 hour time limit between optical camera recalibrations. | Terminated |
| Mar 3, 2011 | Z-2176-2011The anomaly that has been identified with the Optical Guidance Platform (OGP) software may not be always displaying the correct transfer date on the patient file. | Terminated |
| Jan 26, 2009 | Z-0888-2011Potential patient impact -- limit switches for Floorstand may be inadequate for some models and allow for potential impact. | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- May 15, 2007
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files