Modification To: Stockert S5 System

Console, Heart-Lung Machine, Cardiopulmonary Bypass. 510(k) K071318, from Sorin Group Deutschland GmbH (North Attleboro, MA). Product code DTQ, class 2.

As of Nov 21, 2026, K071318 is an FDA 510(k) held by Sorin Group Deutschland GmbH, substantially equivalent on Jul 6, 2007 after 57 days, with 13 recalls tied to it in FDA's recall database.

  • Review time: 57 days
  • Recalls: 13
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The modified S5 and the Pulse Mode Control for the SCP/S5 are used with the SCP Rhythm System in conjunction with the Stöckert S5 System during cardiopulmonary bypass for procedures lasting six (6) hours or less.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K071318, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 28, 2019Z-1119-2020Incorrect labeling; it was identified that some Datapad devices contained incorrect labels. Specifically the manufacturing date in the production identifier was incorrect and the manufacturing date on the unique device…Terminated
Aug 25, 2016Z-2866-2016Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply.Terminated
Dec 4, 2013Z-0856-2014Sorin is initiating a field correction on S5/C5 Heart-lung machine and CP5 centrifugal pump system due to reports of pump speed control knob failures resulting in no response.Terminated
Sep 24, 2013Z-0142-2014Sorin Group is recalling certain models of S5 Perfusion Systems due to malfunction of cardioplegia control.Terminated
Sep 19, 2012Z-0273-2013, Z-0271-2013, Z-0275-2013, Z-0274-2013, Z-0272-2013Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the…Terminated
Feb 23, 2012Z-1326-2012, Z-1328-2012, Z-1327-2012, Z-1325-2012The pump speed control knob is attached to the shaft of an encoder that is used to adjust the speed. Insufficient lubrication within the encoder increases friction over time making it more difficult to turn the pump…Terminated

Other DTQ clearances

The newest 510(k)s for the same product code; their median review was 219 days; see every DTQ device

Record

Type
Special
Received
May 10, 2007
Review panel
Cardiovascular
Regulation
21 CFR 870.4220

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files