NxStage Travel Warmer, Models TW-400 & FWS-500
Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K071263, from Nxstage Medical, Inc. (Lawrence, MA). Product code KDI, class 2.
As of Nov 20, 2026, K071263 is an FDA 510(k) held by Nxstage Medical, Inc., substantially equivalent on Aug 24, 2007 after 109 days, with 1 recall tied to it in FDA's recall database.
- Review time: 109 days
- Recalls: 1
- Cited as a predicate by: 1 device
The NxStage Travel Warmer and Travel Warmer Disposable Set are accessories to the NxStage System One used to warm dialysate prior to administration. E.
- Comfortmate Fluid Warming System, Model # FW-200; Comfortmate Disporsable Disposable Warming Set, Model #FWS-200K012832 · Nxstage Medical, Inc. · Oct 24, 2001
- NxStage Therapy SystemK020858 · Nxstage Medical, Inc. · Apr 17, 2002
- NxStage Express Fluid WarmerK140623 · Nxstage Medical, Inc. · Jun 16, 2014
| Started | Recall | Status |
|---|---|---|
| Jun 11, 2018 | Z-2793-2018There is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is plugged in. The firm has received 7 reports of smoke, sparks, or… | Terminated |
- multiFiltratePRO SystemK261203 · Fresenius Medical Care Renal Therapies Group, LLC · Oct 19, 2026
- Careflux H60;Careflux H70;Careflux V500K261267 · Fresenius Medical Care Renal Therapies Group, LLC · Jul 24, 2026
- NxStage System One with NxViewK252377 · Fresenius Medical Care Renal Therapies Group, LLC · May 15, 2026
- Elisio™-HK260533 · Nipro Medical Corporation · Mar 19, 2026
- multiFlux 130 (F00013123); multiFlux 160 (F00013124)K252459 · Fresenius Medical Care Renal Therapies Group, LLC · Feb 10, 2026
- Tablo Hemodialysis System (PN-0008000, PN-0006000U)K253412 · Outset Medical, Inc. · Jan 26, 2026
- FX CorAL 40; FX CorAL 50K253518 · Fresenius Medical Care Renal Therapies Group, LLC · Dec 18, 2025
- Purema H Hemoconcentrator - PediatricK243920 · Medica USA, Inc. · Sep 19, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 7, 2007
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.5860
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files