Affinity Pediatric Arterial Filter, Model S4014 and With Carmeda Bioactive Surface, Model CB4014

Filter, Blood, Cardiopulmonary Bypass, Arterial Line. 510(k) K071253, from Medtronic Vascular (Minneapolis, MN). Product code DTM, class 2.

As of Nov 19, 2026, K071253 is an FDA 510(k) held by Medtronic Vascular, substantially equivalent on Jul 3, 2007 after 60 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 60 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The AFFINITY® Pediatric Arterial Filter is indicated for use in cardiopulmonary bypass procedures up to 6 hours in duration for the removal of particulate and gaseous microemboli. When Life Depends on Medical Technology AFFINITY Pediatric Arterial Filter Medtronic Confidential Submission Page 8 of 633 Comparison to Predicate Devices The AFFINITY® Pediatric Arterial Filter has the same

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K071253, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 14, 2013Z-1966-2013Medtronic is recalling 148 Affinity Pediatric Arterial Filters from 2 manufacturing lots because a limited quantity of filters may have a small breach in the filter media that may allow unwanted particulate or gaseous…Terminated
May 9, 2008Z-1921-2008Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's…Terminated

Other DTM clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every DTM device

Record

Type
Traditional
Received
May 4, 2007
Review panel
Cardiovascular
Regulation
21 CFR 870.4260

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files