Coaguchek XS Plus System

Test, Time, Prothrombin. 510(k) K071041, from Roche Diagnostics (Indianapolis, IN). Product code GJS, class 2.

As of Nov 18, 2026, K071041 is an FDA 510(k) held by Roche Diagnostics, substantially equivalent on May 11, 2007 after 29 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 29 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The CoaguChek XS Plus System is intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. The system uses fresh capillary or non-anticoagulated venous whole blood.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K071041, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 13, 2018Z-0360-2019Abnormally high INR test results with the affected CoaguChek test stripsTerminated
May 16, 2013Z-1583-2013Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.Terminated
Jul 22, 2009Z-2004-2009The recommended cleaning procedures for the meter may cause a malfunction and falsely elevated test result reports.Terminated

Other GJS clearances

The newest 510(k)s for the same product code; their median review was 299 days; see every GJS device

Record

Type
Special
Received
Apr 12, 2007
Review panel
Hematology
Regulation
21 CFR 864.7750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files