Arca
Skip to content
Tools
Tools

Olympus XBF-UC180F-DT8 Ultrasonic Bronchofibervideoscope Used With Aloka Ssd-Alpha 5/10 Ultrasound System

Ultrasound Bronchoscope. 510(k) K070983, from Olympus Medical Systems Corporation (Center Valley, PA). Product code PSV, class 2.

As of Nov 18, 2026, K070983 is an FDA 510(k) held by Olympus Medical Systems Corporation, substantially equivalent on Jul 5, 2007 after 90 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jul 5, 2007
  • Review time: 90 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The indications for use of OLYMPUS XBF-UC180F-DT8 ULTRASONIC BRONCHOFIBERVIDEOSCOPE used with the ALOKA SSD-Alpha 5/10 ULTRASOUND SYSTEM are as follows: Transesophageal(non-cardiac) Airways and tracheobronchial tree

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K070983, newest first; each opens FDA's recall record
Sep 25, 2023Z-0328-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for…Open, Classified
Jul 27, 2023Z-2494-2023Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.Open, Classified
Aug 8, 2018Z-3304-2018The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.Terminated

Other PSV Clearances

The newest 510(k)s for the same product code; see every PSV device

Record

Type
Traditional
Received
Apr 6, 2007
Review panel
Radiology
Regulation
21 CFR 892.1550

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.