Arca
Skip to content
Tools
Tools

Modification To: Bioknotless BR Anchor / Lupine BR Anchor

Fastener, Fixation, Biodegradable, Soft Tissue. 510(k) K070925, from Depuy Mitek, A Johnson & Johnson Company (Raynham, MA). Product code MAI, class 2.

As of Nov 18, 2026, K070925 is an FDA 510(k) held by Depuy Mitek, A Johnson & Johnson Company, substantially equivalent on May 2, 2007 after 29 days, with 11 recalls tied to it in FDA's recall database.

  • Review time: 29 days
  • Recalls: 11
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Bioknotless Anchor is indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization as follows: OPEN PROCEDURES SHOULDER: 1. Bankart repair 2. SLAP lesion repair; 3. Rotator cuff repair; 4a. Capsule shift/capsulo-labral reconstruction, at the anterior glenoid rim site; 4b.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K070925, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 13, 2023Z-1681-2023, Z-1680-2023, Z-1678-2023, Z-1677-2023, Z-1679-2023IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken…Open
Oct 31, 2011Z-0316-2012, Z-0318-2012, Z-0317-2012, Z-0315-2012, Z-0314-2012Incorrect suture configurationTerminated
Apr 29, 2008Z-1810-2008Lupine BR Anchor w/Orthocord was assembled with an incorrect suture configuration. The suture loop was not assembled at the distal end of the anchor. It was only threaded through the suture hole in the anchor. Risk…Terminated

Other MAI clearances

The newest 510(k)s for the same product code; their median review was 57 days; see every MAI device

Record

Type
Special
Received
Apr 3, 2007
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.