Arca
Skip to content
Tools
Tools

Gambro Cartridge Blood Sets

Set, Tubing, Blood, With And Without Anti-Regurgitation Valve. 510(k) K070414, from Gambro Renal Products (Lakewood, CO). Product code FJK, class 2.

As of Nov 20, 2026, K070414 is an FDA 510(k) held by Gambro Renal Products, substantially equivalent on Oct 12, 2007 after 242 days, with 12 recalls tied to it in FDA's recall database.

  • Review time: 242 days
  • Recalls: 12
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Gambro Cartridge® Blood Set is intended for single use in a hemodialysis treatment using the Phoenix® and Centrysystem® 3 Dialysis Delivery Systems. The Low Weight model is used when a low extra-corporeal blood volume is recommended. The Low Weight model with a priming volume of 75 ml is indicated for patients with body weight greater than 20 Kg and lower or equal to 40 Kg.

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K070414, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 14, 2021Z-1684-2021There is a potential leak between the venous patient connector and patient's hemodialysis access.Terminated
Dec 30, 2020Z-1120-2021, Z-1119-2021, Z-1121-2021, Z-1118-2021, Z-1122-2021complaints of kinked access lines observed during treatment using cartridge sets for hemodialysis treatments were received.Open
Apr 17, 2020Z-2196-2020, Z-2194-2020, Z-2193-2020, Z-2195-2020Potential disconnection of tubing set.Terminated
May 2, 2008Z-1860-2008, Z-1861-2008Dialysis tubing sets may have occlusions restricting blood flow.Terminated

Other FJK clearances

The newest 510(k)s for the same product code; their median review was 261 days; see every FJK device

Record

Type
Traditional
Received
Feb 12, 2007
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5820

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files