Arthrocare Ent Plasma Wand
Electrosurgical, Cutting & Coagulation & Accessories. 510(k) K070374, from Arthrocare Corp. (Sunnyvale, CA). Product code GEI, class 2.
As of Nov 18, 2026, K070374 is an FDA 510(k) held by Arthrocare Corp., substantially equivalent on Apr 25, 2007 after 76 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Apr 25, 2007
- Review time: 76 days
- Recalls: 1
- Cited as a predicate by: 9 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe ArthroCare ENT Plasma Wands is for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including: • Adenoidectomy • Cysts • Head, Neck, Oral, and Sinus Surgery • Mastoidectomy • Myringotomy with Effective Hemorrhage Control • Nasal Airway Obstruction by Reduction of Hypertrophic Nasal Turbinates • Nasopharyngeal/Laryngeal…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Entec Plasma WandsK014290 · Arthrocare Corp. · Mar 28, 2002
- Arthrocare Controller (System 2000 And 8000), Arthrocare Patient Cable, Foot Control, Power Cord And WandsK063538 · Arthrocare Corp. · Dec 1, 2006
Cited as a Predicate
9 later 510(k) summaries name this device as a predicate- ENT Plasma Wands, Turbinator Wand, Werewolf Irrigation Tube Set, Werewolf ENT Adapter, Werewolf Coblation SystemK202006 · ArthroCare Corporation · Dec 18, 2020
- WEREWOLF COBLATION System, COBLATION HALO WandK192027 · ArthroCare Corporation · Dec 20, 2019
- Reprocessed ArthroCare ENT CoblatorK172647 · Renovo, Inc. · Feb 21, 2018
- EVac 70 Xtra Plasma Wand with Integrated Cable, PROcise XP Plasma Wand with Integrated CableK142999 · ArthroCare Corporation · Nov 19, 2014
- Ciq Controller, Ciq WandsK123353 · ArthroCare Corporation · Apr 5, 2013
- Arthrocare Turbinator WandK122652 · ArthroCare Corporation · Jul 2, 2013
- Harmonic Focus ShearsK100597 · Ethicon Endo-Surgery, LLC · Apr 7, 2010
- Medississ Reprocessedent Ablation Wands (Coblators), MODEL#EIC8872-01K100543 · Medisiss · Jul 23, 2010
- Peak Plasmablade Tna Tonsil And Adenoid Tissue Dissection DeviceK083415 · Peak Surgical, Inc. · Jun 5, 2009
Recalls
Recalls FDA ties to K070374, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Jun 2, 2015 | Z-2105-2015Wands inspected with the adaptor fixture were released to the field not meeting the dielectric (Hipot) specifications. | Terminated |
Other GEI Clearances
The newest 510(k)s for the same product code; their median review was 167 days; see every GEI device- SP Electrocautery Device (SP18 Max);SP Electrocautery Device (SP20 Max)K262720 · Single Pass, Inc. · Oct 26, 2026
- SafeAir Electrosurgical Generator with Smoke Evacuation (0795-001-000);SafeAir Generator 3-Pedal Footswitch…K262386 · Stryker Instruments · Oct 20, 2026
- HARMONIC 1100 Shears (HAR1120); HARMONIC 1100 Shears (HAR1136)K254196 · Ethicon Endo-Surgery, LLC · Sep 17, 2026
- Neptune® SafeAir® 10ft Smoke Evacuation Pencil, Square Coated, 70mm Blade, Push Button Switch (0703-046-100);Neptune®…K261959 · Stryker Instruments · Sep 17, 2026
- LigaSure RAS Maryland Jaw Vessel Sealer/Divider, Nano-coatedK261070 · Covidien, LLC · Sep 15, 2026
- Potenza PrimeK254185 · Jeisys Medical Incorporated · Sep 11, 2026
- Ellian Enigma RF DeviceK262364 · Ellian Corp · Sep 2, 2026
- Scarlet Pro (Scarlet Pro)K253859 · VIOL Co., Ltd. · Aug 27, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 8, 2007
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4400
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files