Arca
Skip to content
Tools
Tools

Arthrocare Ent Plasma Wand

Electrosurgical, Cutting & Coagulation & Accessories. 510(k) K070374, from Arthrocare Corp. (Sunnyvale, CA). Product code GEI, class 2.

As of Nov 18, 2026, K070374 is an FDA 510(k) held by Arthrocare Corp., substantially equivalent on Apr 25, 2007 after 76 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Apr 25, 2007
  • Review time: 76 days
  • Recalls: 1
  • Cited as a predicate by: 9 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The ArthroCare ENT Plasma Wands is for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including: • Adenoidectomy • Cysts • Head, Neck, Oral, and Sinus Surgery • Mastoidectomy • Myringotomy with Effective Hemorrhage Control • Nasal Airway Obstruction by Reduction of Hypertrophic Nasal Turbinates • Nasopharyngeal/Laryngeal…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

9 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K070374, newest first; each opens FDA's recall record
StartedReasonStatus
Jun 2, 2015Z-2105-2015Wands inspected with the adaptor fixture were released to the field not meeting the dielectric (Hipot) specifications.Terminated

Other GEI Clearances

The newest 510(k)s for the same product code; their median review was 167 days; see every GEI device

Record

Type
Traditional
Received
Feb 8, 2007
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.4400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files