Fresenius 2008K @Home With Wireless Wetness Detector
Hemodialysis System For Home Use. 510(k) K070049, from Fresenius Medical Care North America (Waltham, MA). Product code ONW, class 2.
As of Nov 19, 2026, K070049 is an FDA 510(k) held by Fresenius Medical Care North America, substantially equivalent on Feb 3, 2011 after 1,491 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 1,491 days
- Recalls: 7
- Cited as a predicate by: 1 device
The Wireless Wetness Detector is indicated for use with the Fresenius 2008K@home hemodialysis machine and is an optional accessory to aid in the detection of blood and water leaks during hemodialysis. Home hemodialysis using the detector must be observed by a trained and qualified person as prescribed by a physician. 551 K070049 892393 F.
- Fresenius 2008KK994267 · Fresenius Medical Care · Mar 16, 2000
- Fresenius 2008K Home Hemodialysis Machine With Wireless Wetness Detector (Wetalert)K121421 · Fresenius Medical Care North America, Design Cente · Jan 17, 2013
| Feb 20, 2015 | Z-1549-2015The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen. | Terminated |
| Nov 12, 2013 | Z-0629-20142008 Series Hemodialysis Machines complaints of saline bags inappropriately filling during recirculation and setup | Terminated |
| Aug 29, 2013 | Z-0185-2014, Z-0184-2014, Z-0186-2014Updated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss | Terminated |
| Aug 11, 2011 | Z-3269-2011Blood lines coiling tabs #1 and #5 on the arterial line are not in the correct positions | Terminated |
| May 4, 2011 | Z-3083-2011Upgrade software to implement new on screen graphics and instructions, the addition of a bracket on the side of the machine to temporarily hold the bloodlines during set up in order to conform to approved 510(k) | Terminated |
- SC+ Hemodialysis Device (SC-14269); Dialysate Cartridge (SC-14656); Blood Tube Set (SC-14651)K242269 · Quanta Dialysis Technologies, Ltd. · Nov 1, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 4, 2007
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.5860
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files