Arca
Skip to content
Tools
Tools

Power Injectable Implantable Infusion Port With Silicone Lumen (Attachable and Pre-Attached)

Port & Catheter, Implanted, Subcutaneous, Intravascular. 510(k) K070003, from Medcomp (Harleysville, PA). Product code LJT, class 2.

As of Nov 22, 2026, K070003 is an FDA 510(k) held by Medcomp, substantially equivalent on May 15, 2007 after 132 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 132 days
  • Recalls: 3
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

THE POWER INJECTABLE IMPLANTABLE INFUSION PORT IS INDICATED FOR PATIENT THERAPIES REQUIRING REPEATED ACCESS TO THE VASCULAR SYSTEM. THE PORT SYSTEM CAN BE USED FOR INFUSION OF MEDICATIONS, I.V. FLUIDS, PARENTERAL NUTRITION SOLUTIONS, BLOOD PRODUCTS AND FOR THE WITHDRAWAL OF BLOOD SAMPLES.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K070003, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 12, 2012Z-1626-201250 trays of Dignity Low Profile CT Port were labeled as Dignity Mid-Sized CT Port.Terminated
Apr 2, 2012Z-1868-2012Product was packaged with the incorrect port and does not match the label. The kit is labeled as an 8F Low Profile Pro-Fuse CT Port. The kit contains an 8F Midsize Dignity CT Port.Terminated
Jul 2, 2010Z-2235-2010Recalled product was packaged with the incorrect port and does not match the label. The kit is labeled as a 6.6F "Dignity" Low Profile CT Port. The kit contains a 6.6F "Pro-Fuse" Low Profile CT Port.Terminated

Other LJT clearances

The newest 510(k)s for the same product code; their median review was 86 days; see every LJT device

Record

Type
Traditional
Received
Jan 3, 2007
Review panel
General Hospital
Regulation
21 CFR 880.5965

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files