Arca
Skip to content
Tools
Tools

Variant Ii Turbo Hemoglobin A1C Program, Hemoglobin Testing System With Cdm 4.0

Assay, Glycosylated Hemoglobin. 510(k) K063400, from Bio-Rad Laboratories, Inc. (Hercules, CA). Product code LCP, class 2.

As of Nov 20, 2026, K063400 is an FDA 510(k) held by Bio-Rad Laboratories, Inc., substantially equivalent on Dec 1, 2006 after 22 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 22 days
  • Recalls: 4

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Bio-Rad VARIANT II TURBO Hemoglobin A1C Program with CDM 4.0 is intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The Bio-Rad VARIANT II TURBO Hemoglobin A1C Program with CDM 4.0 is intended for Professional Use Only. For In Vitro Diagnostic Use.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K063400, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 24, 2010Z-1342-2011The firm received customer complaints associated with the Update Kit CD-ROM not being able to generate a valid calibration after being uploaded, resulting in patient results not being generated.Terminated
Dec 8, 2009Z-1898-2011Wrong lot number was found on the Hemoglobin Cartridge Insert . The insert did not match the lot number to the device cartridge or the device cartridge label or on the Update Kit CD.Terminated
Oct 12, 2009Z-0867-2010Product update kit CD rom will not upload. As a result, the device cannot be run.Terminated
Oct 17, 2008Z-1041-2009Sample results may be assigned to the wrong barcode number and tube position numbers, leading to discrepant sample results.Terminated

Other LCP clearances

The newest 510(k)s for the same product code; see every LCP device

Record

Type
Special
Received
Nov 9, 2006
Review panel
Hematology
Regulation
21 CFR 864.7470

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files