Definium 5000 X-Ray System
Solid State X-Ray Imager (Flat Panel/Digital Imager). 510(k) K063283, from Sedecal., Sa. (Deer Field, IL). Product code MQB, class 2.
As of Nov 18, 2026, K063283 is an FDA 510(k) held by Sedecal., Sa., substantially equivalent on Nov 17, 2006 after 18 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 18 days
- Recalls: 3
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Definium™ 5000 X-Ray System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Sedecal X Plus LP PlusK062335 · Sedecal., Sa. · Sep 14, 2006
- Ge Tethered Portable Digital Radiographic DetectorK041922 · General Electric Co. · Jul 28, 2004
- Urs LP Digital, Millennium, RadproK042876 · Sedecal., Sa. · Nov 18, 2004
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Perkinelmer Xrd 1622 AP3 Med Flat Panel DetectorK122495 · Perkinelmer, Inc. · Nov 28, 2012
Recalls
Recalls FDA ties to K063283, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Mar 24, 2009 | Z-1118-2009A defect in the AEC board in the system's control causes failure to terminate additional exposures at the exposure time selected for the initial exposure. | Terminated |
| Dec 17, 2007 | Z-1133-2008, Z-1134-2008GE Definium 5000 Collimator failed compliance testing due to a blade sizing issue. | Terminated |
Other MQB clearances
The newest 510(k)s for the same product code; their median review was 179 days; see every MQB device- 0909FCE, 0909FCE-HSK253800 · Rayence Co., Ltd. · Jul 31, 2026
- Vivix-S 4386W (FXRD-4386WA)K260820 · Vieworks Co., Ltd. · Jun 18, 2026
- Universal 1417PIK252996 · Konica Minolta Healthcare Americas, Inc. · Jun 12, 2026
- Lux HD 2530 detector (Lux HD 2530)K252911 · Iray Imaging Technology (Haining) Limited · Oct 9, 2025
- Yushan X-Ray Flat Panel DetectorK250211 · Innocare Optoelectronics Corp. · Jul 22, 2025
- Skr 3000K250665 · Konica Minolta, Inc. · Jun 17, 2025
- Wireless/ Wired X-Ray Flat Panel DetectorsK243734 · Allengers Medical Systems Limited · Apr 18, 2025
- Expd 114; Expd 114G; Expd 114P; Expd 114PGK242770 · DRTECH Corporation · Mar 20, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Oct 30, 2006
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files