System, Image Processing, Radiological. 510(k) K063267, from Stentor Inc., A Philips Medical Systems Co. (Brisbane, CA). Product code LLZ, class 2.

As of Nov 18, 2026, K063267 is an FDA 510(k) held by Stentor Inc., A Philips Medical Systems Co., substantially equivalent on Nov 9, 2006 after 10 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 10 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

iSite PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K063267, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 23, 2011Z-1881-2011A software anomaly that occurs when clinicians use Volume Vision to view images in 3D and uses the measurement tool to measure regions of interest. This could lead to an incorrect interpretation of the measurements…Terminated
Dec 29, 2008Z-0948-2009Product may display clinical reports with the study date and time in GMT format, which may not be obvious to the end user, who may expect to see the local time in the report.Terminated
Aug 29, 2008Z-0118-2009Two defects have been identified: 1) potential to display a patient on the canvas page that is different from the patient whose images are displayed on diagnostic monitors when using Conference Presentation States and…Terminated
Jan 29, 2008Z-1444-2008When using the Freehand Region of Interest (ROI) tool, errors may occur in area calculation yielding incorrect results. The firm discovered that a descrepancy exists in the measurement provided by the tool.Terminated

Other LLZ clearances

The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device

Record

Type
Traditional, third-party review
Received
Oct 30, 2006
Review panel
Radiology
Regulation
21 CFR 892.2050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files