Monarch Iii Iol Delivery System
Lens, Guide, Intraocular. 510(k) K063155, from Alcon Manufacturing, Ltd. (Fort Worth, TX). Product code KYB, class 1.
As of Nov 18, 2026, K063155 is an FDA 510(k) held by Alcon Manufacturing, Ltd., substantially equivalent on Mar 27, 2007 after 161 days, with 1 recall tied to it in FDA's recall database.
- Review time: 161 days
- Recalls: 1
- Cited as a predicate by: 6 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe MONARCH® III IOL Delivery System is used for implantation of ALCON® qualified ACRYSOF® intraocular lenses into the eye following cataract removal. 4. Brief Summary of Nonclinical Test and Results: Alcon has provided a Statement of Conformance (Attachment J) that the Monarch® III IOL Delivery System Cartridge D and Handpiece H4 intended to be introduced, will conform in all respects with the…
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Monarch Ii Iol Delivery SystemK003768 · Alcon Research, Ltd. · Jan 5, 2001
- Monarch Ii Iol Delivery SystemK001157 · Alcon Research, Ltd. · Jun 27, 2000
Cited as a predicate
6 later 510(k) summaries name this device as a predicate- BL-Cart IOL Delivery CartridgeK182965 · Ast Products, Inc. · Mar 29, 2019
- Bausch & Lomb Injector SystemK131958 · Bausch & Lomb · Jan 9, 2014
- Infiniti(R) Vision SystemK120912 · Alcon Manufacturing, Ltd. · Jun 15, 2012
- Bausch & Lomb Iol InjectorK113852 · Bausch & Lomb, Inc. · Sep 28, 2012
- Monarch Iii Iol Delivery System, Monarch Iii C Cartridge, Monarch Iii Handpiece (H4)K112977 · Alcon Research, Ltd. · Mar 27, 2012
- Monarch Iii Iol Delivery System; Monarch Iii P Catridge; Monarch Iii Handpiece (H4)K101774 · Alcon Manufacturing, Ltd. · Oct 13, 2011
Recalls
Recalls FDA ties to K063155, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Started | Recall | Status |
|---|---|---|
| Nov 13, 2024 | Z-0835-2025Increase in complaints associated with a molding defect within the inner lumen of delivery cartridges which could result in a scratch on the intraocular lens (IOL) upon delivery. | Open |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 17, 2006
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.4300
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files