Arca
Skip to content
Tools
Tools

Deltec Cozmo Insulin Infusion Pump With Cozmonitor Blood Glucose Meter, Models 21-1815, 21-1816, 21-1817

Pump, Infusion, Insulin. 510(k) K062323, from Smiths Medical MD, Inc. (St. Paul, MN). Product code LZG, class 2.

As of Nov 20, 2026, K062323 is an FDA 510(k) held by Smiths Medical MD, Inc., substantially equivalent on Nov 1, 2006 after 84 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 84 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Deltec Cozmo® Insulin Pump The Deltec Cozmo ® Insulin Pump is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes.

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K062323, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 20, 2008Z-0989-2009, Z-0988-2009Smiths Medical has become aware of a display irregularity with the Deltec Cozmo Insulin Pump, Model 1800 and Model 21-1700 containing Model 1800 software. There have been adverse events reported that the amount of…Terminated
Mar 10, 2008Z-1623-2008, Z-1622-2008Smith Medical discovered an issue with a motor component in specific serial numbers of the Deltec Cozmo Insulin Pump, Model Numbers 1700 and 1800. This issue affects how the motor operates and may cause an over-delivery…Terminated
Feb 22, 2008Z-1591-2008Smith Medical has received reports of Deltec Cozmo Insulin Pumps powering down without an alarm. Investigation concluded that if the battery cap is not fully tightened, the Pump may power down and a brief "chirp" will…Terminated

Other LZG clearances

The newest 510(k)s for the same product code; their median review was 146 days; see every LZG device

Record

Type
Special
Received
Aug 9, 2006
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files