Mantis Spinal System
Orthosis, Spondylolisthesis Spinal Fixation. 510(k) K061813, from Stryker Spine (Allendale, NJ). Product code MNH, class 2.
As of Nov 17, 2026, K061813 is an FDA 510(k) held by Stryker Spine, substantially equivalent on Sep 19, 2006 after 84 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 19, 2006
- Review time: 84 days
- Recalls: 2
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe MANTISTM Spinal System is intended for percutaneous, posterior, non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Modification To Xia Spinal SystemK043473 · Stryker Spine · Jan 12, 2005
- Xia Spinal System, Xia 4.5 Spinal SystemK060361 · Stryker Spine · Mar 6, 2006
- Xia Spine SystemK002858 · Howmedica Osteonics Corp. · Oct 13, 2000
- Xia Spinal SystemK013823 · Howmedica Osteonics Corp. · Dec 19, 2001
- Osteonics Spinal SystemK951725 · Osteonics Corp. · Sep 8, 1995
- Xia Spine SystemK984251 · Howmedica, Inc. · Feb 24, 1999
- Cd Horizon Spinal SystemK001255 · Medtronic Sofamor Danek, Inc. · Jul 11, 2000
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- Mantis & Mantis Redux Spinal System, Radius Spinal System, Trio & Trio+ Spinal Fixation…K133188 · Stryker Spine · Dec 17, 2013
- Picasso Mis Spinal Fixation SystemK120714 · Accel Spine · Aug 9, 2012
- Focus Mis SystemK112643 · L&K BIOMED Co., Ltd. · Jan 4, 2012
- Mantis Spinal SystemK073151 · Stryker Corp. · Dec 6, 2007
Recalls
Recalls FDA ties to K061813, newest first; each opens FDA's recall record| Apr 30, 2009 | Z-1611-2009The Mantis Rod Inserter Shaft was found to have deformation of the stainless stell split sphere when subjected to high forces, prohibiting the user from being able to grip the rod implantation and the split sphere may… | Terminated |
| Apr 30, 2009 | Z-1612-2009The Mantis Rod Inserter Shaft was found to have deformation of the stainless stell split sphere when subjected to high forces, prohibiting the user from being able to grip the rod implantation and the split sphere may… | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 27, 2006
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3070
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files