Boehringer Laboratories Suction Pump System
Negative Pressure Wound Therapy Powered Suction Pump. 510(k) K061788, from Boehringer Laboratories (Norristown, PA). Product code OMP, class 2.
As of Nov 18, 2026, K061788 is an FDA 510(k) held by Boehringer Laboratories, substantially equivalent on Jul 11, 2006 after 15 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 15 days
- Recalls: 5
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Boehringer Laboratories Suction Pump System is intended for the application of suction (negative pressure) to wounds to promote wound healing and for the removal of fluids, including wound exudate, irrigation fluids, body fluids and infectious materials.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Boehringer Laboratories Suction Pump SystemK060277 · Boehringer Laboratories · Mar 3, 2006
- Versatile 1 Wound Vacuum SystemK042134 · Blue Sky Medical Group Incorporated · Nov 1, 2004
- Modification To Versatile 1 Wound Vacuum SystemK052456 · Blue Sky Medical Group Incorporated · Nov 4, 2005
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Venturi Negative Pressure Wound Therapy SystemK080897 · Talley Medical · Jul 15, 2008
Recalls
Recalls FDA ties to K061788, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jul 9, 2010 | Z-2609-2010Reduced vacuum capacity has been reported with a small number of units caused by electrostatic discharge (ESD) or static electricity. | Terminated |
| Jul 29, 2009 | Z-2044-2009, Z-2043-2009, Z-2045-2009Labeling: The expiration date does not appear on the product label. | Terminated |
| Jul 13, 2009 | Z-1814-2009Labeling: the actual manufacturing date (2/2009) and actual expiration date (2/2011) are switched. | Terminated |
Other OMP clearances
The newest 510(k)s for the same product code; their median review was 240 days; see every OMP device- Wound VOX NPWT System (V1-01-01); Wound VOX 300ml Collection Canister & Supply Tubing (CC-00-03); Wound VOX Y-Connector…K262232 · Bechtel Medical, Inc. · Nov 6, 2026
- RENASYS-G 15Fr Channel Drain Gauze Dressing Kit (66802257); RENASYS 15Fr Channel Drain Accessory Kit (66802253)K260306 · Smith & Nephew Medical, Ltd. · Oct 30, 2026
- RENASYS ABDO SHIELD Temporary Abdominal Closure Kit (66807747)K261982 · Smith & Nephew Medical, Ltd. · Oct 23, 2026
- MeCare Alpha (MCA2019)K260426 · IVT Medical, Ltd. · Oct 8, 2026
- RENASYS DYNAFLOW Kit (66807741)K261338 · Smith & Nephew Medical, Ltd. · Sep 11, 2026
- ConvaVAC™ Negative Pressure Wound Therapy SystemK253263 · Convatec Limited, Gdc Building · Jun 24, 2026
- Curasys 2 (EMDN0003)K253199 · CG Bio Co., Ltd. · Jun 23, 2026
- Renasys Edge (66803126)K260646 · Smith & Nephew Medical, Ltd. · May 27, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Jun 26, 2006
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4780
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files