Iled
Light, Surgical, Ceiling Mounted. 510(k) K061317, from Trumpf Kreuzer Medizin Systeme GmbH + Co.Kg (Munchen, DE). Product code FSY, class 2.
As of Nov 28, 2026, K061317 is an FDA 510(k) held by Trumpf Kreuzer Medizin Systeme GmbH + Co.Kg, substantially equivalent on Jun 22, 2006 after 42 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 42 days
- Recalls: 7
- Cited as a predicate by: 5 devices
The surgical light iLED is intended to illuminate locally the operating site on the patient's body with a high intensity, shadow fee, "cold" light.
- Polaris 100/200K123776 · Drager Medical GmbH · Mar 8, 2013
- 21ST Century Centurion Exceled and System Two Led Lighting SystemsK100388 · Medical Illumination International, Inc. · Jun 22, 2010
- Mach Led MCK093010 · Dr. Mach GmbH & Co. KG · Apr 1, 2010
- Nuvo Verde Led Lighting SystemK083323 · Nuvo, Inc. · Jan 28, 2009
- Harmony LED-1 Surgical Lighting SystemK072072 · STERIS Corporation · Oct 5, 2007
| Nov 2, 2015 | Z-0563-2016The firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head… | Completed |
| Jul 27, 2015 | Z-2353-2015, Z-2357-2015, Z-2354-2015, Z-2356-2015Complaints received of the front joint of the spring arm AC 2000 could break at the interface to the monitor. | Terminated |
| May 29, 2012 | Z-2311-2012Cracks can arise on the welded seam on the rear joint. | Terminated |
| Feb 27, 2009 | Z-1122-2009Cracks may form around the screw connections of the cover plates causing brittleness, which may lead to fragments breaking off and falling into the operating area. | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- May 11, 2006
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4580
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files