Elia Ccp, Model 14-5515-01 and 1 more
Antibodies, Anti-Cyclic Citrullinated Peptide (Ccp). 510(k) K061165, from Sweden Diagnostics (Us), Inc. (Portage, MI). Product code NHX, class 2.
As of Nov 25, 2026, K061165 is an FDA 510(k) held by Sweden Diagnostics (Us), Inc., substantially equivalent on Aug 3, 2006 after 99 days, with 1 recall tied to it in FDA's recall database.
- Review time: 99 days
- Recalls: 1
- Cited as a predicate by: 1 device
EliA CCP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to cyclic citrullinated peptides (CCP) in human serum and plasma. The presence of anti-CCP antibodies can be used in conjunction with clinical findings and other laboratory tests as an aid in the clinical diagnosis of rheumatoid arthritis (RA).
- Diastat Anti Cyclic Citrullinated Peptide (Anti-Ccp) ElisaK023285 · Axis-Shield Diagnostics, Ltd. · Oct 9, 2002
- EliA anti-TG Immunoassay, EliA anti-TPO Immunoassay, EliA Thyroid Positive Control 250, EliA Thyroid Positive Control…K151799 · Phadia AB · Mar 25, 2016
| Started | Recall | Status |
|---|---|---|
| Jan 30, 2017 | Z-1253-2017Erroneous EliA test results related to problems with liquid level detection and diluent pipetting in Phadia 250 instruments, and deformed bottles. Investigation showed problems with the liquid level detection in diluent… | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 26, 2006
- Review panel
- Immunology
- Regulation
- 21 CFR 866.5775
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files