Arca
Skip to content
Tools
Tools

Dimension Vista Albumin, Alkaline Phosphatase, Aspartate Amino Transferase, Carbamazepine & Alanine Amino Transferase

Nadh Oxidation/Nad Reduction, Alt/Sgpt. 510(k) K061020, from Dade Behring, Inc. (Newark, DE). Product code CKA, class 1.

As of Nov 20, 2026, K061020 is an FDA 510(k) held by Dade Behring, Inc., substantially equivalent on May 10, 2006 after 27 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 27 days
  • Recalls: 1

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Dimension Vista™ Albumin Flex® reagent cartridge (ALB) is a device intended to measure the albumin concentration in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K061020, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jun 29, 2017Z-1657-2018Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.Terminated

Record

Type
Special
Received
Apr 13, 2006
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files