Genius 2 Infrared Tympanic Electronic Thermometer, Model 303000

Continuous Measurement Thermometer. 510(k) K060649, from Tyco Healthcare Group LP (Mansfield, MA). Product code FLL, class 2.

As of Nov 26, 2026, K060649 is an FDA 510(k) held by Tyco Healthcare Group LP, substantially equivalent on May 22, 2006 after 70 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 70 days
  • Recalls: 5

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Kendall Genius 2 Infrared Tympanic Electronic Thermometer is intended for use in patients in Acute and Alternate care settings to provide temperature measurements from the tympanic membrane and equivalent measurements of oral, core, rectal, and axillary temperature based on the tympanic reading.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K060649, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Oct 28, 2019Z-0992-2020, Z-0991-2020, Z-0994-2020, Z-0990-2020, Z-0993-2020The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( 0.2C for Genius 2 and 0.3C for Genius…Terminated

Other FLL clearances

The newest 510(k)s for the same product code; their median review was 207 days; see every FLL device

Record

Type
Traditional
Received
Mar 13, 2006
Review panel
General Hospital
Regulation
21 CFR 880.2910

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files