Arca
Skip to content
Tools
Tools

Css Cannulated Screw System

Screw, Fixation, Bone. 510(k) K060428, from Orthohelix Surgical Designs, Inc. (Akron, OH). Product code HWC, class 2.

As of Nov 18, 2026, K060428 is an FDA 510(k) held by Orthohelix Surgical Designs, Inc., substantially equivalent on Mar 7, 2006 after 14 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 14 days
  • Recalls: 4
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The CSS Cannulated Screw System will be indicated for use in long and small bone fracture fixation, which includes the following; • Fractures of the tarsals and metatarsals • Fractures of the olecranon, distal humerous • Fractures of the radius and ulna • Patella fractures • Distal tibia and pilon fractures • Fractures of the fibula, medial malleolus, os calcis • Tarso-metatarsal and metatarsal…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K060428, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 15, 2010Z-0468-2011, Z-0466-2011, Z-0467-2011The print for modified CSS depth gauges has inaccurate definition of the etched screw length. The error results in a length reading that is longer (5 mm in the case of the 7.0 and 4mm in the cases of the 5.5 and 4.0)…Terminated
Apr 7, 2009Z-1250-2010A pinhead sized disc was found in the cannulation of a MSD-010-55-60S screw.Terminated

Other HWC clearances

The newest 510(k)s for the same product code; their median review was 157 days; see every HWC device

Record

Type
Traditional
Received
Feb 21, 2006
Review panel
Orthopedic
Regulation
21 CFR 888.3040

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files