Confidence High Viscosity Bone Cement
Cement, Bone, Vertebroplasty. 510(k) K060300, from Disc-O-Tech Medical Technologies, Ltd. (Herzliya, IL). Product code NDN, class 2.
As of Nov 17, 2026, K060300 is an FDA 510(k) held by Disc-O-Tech Medical Technologies, Ltd., substantially equivalent on Jun 21, 2006 after 135 days, with 8 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jun 21, 2006
- Review time: 135 days
- Recalls: 8
- Cited as a predicate by: 7 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Kyphx Hv-R Bone Cement, Model C01AK041584 · Kyphon, Inc. · Jul 7, 2004
- Simplex XK032945 · Stryker Corp. · May 25, 2004
- Vertefix Vertebroplasty Procedure SetK042691 · Cook, Inc. · Nov 8, 2005
Cited as a Predicate
7 later 510(k) summaries name this device as a predicate- CONFIDENCE Spinal Cement SystemK201831 · Depuy Spine, Inc. · Aug 31, 2020
- Confidence High Viscosity Spinal Cement, 16G / 20GK112907 · Medos International SARL · Dec 22, 2011
- Kyphon Xpede Bone CementK102397 · Medtronic · Feb 28, 2011
- Cortoss Bone Augmentation MaterialK080108 · Orthovita, Inc. · Jun 5, 2009
- Mesh Fenestrated Introducer Needle (Mesh Introducer Needle)K071375 · Disc-O-Tech Medical Technologies, Ltd. · Dec 21, 2007
- Confidence Fenestrated Introducer NeedleK063067 · Disc-O-Tech Medical Technologies, Ltd. · Nov 3, 2006
- Confidence Ex High Viscosity Bone CementK062424 · Disc-O-Tech Medical Technologies, Ltd. · Sep 14, 2006
Recalls
Recalls FDA ties to K060300, newest first; each opens FDA's recall record| Dec 15, 2017 | Z-1300-2018Certain lots of the spinal cement may contain incorrect amounts of contrast medium so that the product is difficult or impossible to visualize on x-ray. | Terminated |
| Apr 17, 2013 | Z-2134-2013During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement. | Terminated |
| Apr 17, 2013 | Z-2132-2013During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement. | Terminated |
| Apr 17, 2013 | Z-2133-2013During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement. | Terminated |
| Jan 15, 2013 | Z-0963-2013Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck | Terminated |
| Jan 15, 2013 | Z-0960-2013Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck | Terminated |
| Jan 15, 2013 | Z-0962-2013Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck | Terminated |
| Jan 15, 2013 | Z-0961-2013Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck | Terminated |
Other NDN Clearances
The newest 510(k)s for the same product code; their median review was 168 days; see every NDN device- XeliteMed SuperM-Fix Spinal Bone CementK251896 · Xelite Biomed , Ltd. · Jan 28, 2026
- VCFix Spinal SystemK250637 · Amber Implants · May 30, 2025
- XeliteMed VertehighFix High Viscosity Spinal Bone Cement SystemK243537 · Xelite Biomed , Ltd. · Apr 11, 2025
- XeliteMed VertehighFix High Viscosity Spinal Bone CementK241775 · Xelite Biomed , Ltd. · Sep 18, 2024
- Balloon Inflation SystemK232842 · Ningbo Hicren Biotechnology Co., Ltd. · Feb 29, 2024
- V-STRUT® Vertebral ImplantK240084 · Hyprevention · Feb 9, 2024
- Renova Spine Balloon CatheterK231340 · Biopsybell S.R.L. · Oct 30, 2023
- Injection Pin (KIP(02031-02061) (03031-03061))K221697 · Slk Ortho, LLC · Mar 3, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 6, 2006
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3027
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files