Arca
Skip to content
Tools
Tools

Preformance Temporary Cylinder

Abutment, Implant, Dental, Endosseous. 510(k) K060291, from Implant Innovations, Inc. (Palm Beach Gardens, FL). Product code NHA, class 2.

As of Nov 21, 2026, K060291 is an FDA 510(k) held by Implant Innovations, Inc., substantially equivalent on Feb 27, 2006 after 21 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 21 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The PreFormance Temporary Cylinders are intended for use as an accessory to endosseous dental implants to support a prosthetic device in a partially or fully edentulous patient. They are intended for use to support single and multiple unit prostheses in the mandible or maxilla for up to 180 days during endosseous and gingival healing, and are for non occlusal loading of single and multiple unit…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K060291, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 25, 2015Z-2415-2016, Z-2493-2016Pouches may not have been sealed during packing.Terminated
Oct 8, 2007Z-0128-2008Incorrect assembly: The product does not allow for the screw to pass through the access hole and engage the implant properly. Therefore, the provisional restoration will not be seated on the implant platform.Terminated

Other NHA clearances

The newest 510(k)s for the same product code; their median review was 169 days; see every NHA device

Record

Type
Special
Received
Feb 6, 2006
Review panel
Dental
Regulation
21 CFR 872.3630

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files