Dimension Iron Calibrator
Calibrator, Primary. 510(k) K060266, from Dade Behring, Inc. (Newark,, DE). Product code JIS, class 2.
As of Nov 20, 2026, K060266 is an FDA 510(k) held by Dade Behring, Inc., substantially equivalent on Mar 9, 2006 after 36 days, with 1 recall tied to it in FDA's recall database.
- Review time: 36 days
- Recalls: 1
- Cited as a predicate by: 1 device
The IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the IRON method for the Dimension® clinical chemistry system.
- Dimension(R) XL Clinical Chemistry SystemK944093 · E.I. Dupont DE Nemours & Co., Inc. · Nov 10, 1994
- Dimension Vista Iron Flex Reagent Cartridge and CalibratorK061793 · Dade Behring, Inc. · Aug 7, 2006
| Started | Recall | Status |
|---|---|---|
| Dec 20, 2012 | Z-0745-2013Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may… | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 1, 2006
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1150
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files