Arca
Skip to content
Tools
Tools

Olympus Guide Sheath, Model XBO1-836-13

Bronchoscope (Flexible Or Rigid). 510(k) K060243, from Olympus Medical Systems Corporation (Melville, NY). Product code EOQ, class 2.

As of Nov 19, 2026, K060243 is an FDA 510(k) held by Olympus Medical Systems Corporation, substantially equivalent on Jun 23, 2006 after 143 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 143 days
  • Recalls: 9
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

This instrument has been designed to be used with Olympus bronchosscopes, endo-therapy accessories and ultrasound probe to guide the endo-therapy accessories or ultrasound probe to the target area within the respiratory organs.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K060243, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 15, 2025Z-1071-2025Potential for the radiopaque tip of the guide sheath component to disassociate and fall off into the patient.Open
Jan 4, 2021Z-0970-2021, Z-0972-2021, Z-0971-2021, Z-1053-2021, Z-1050-2021, Z-1052-2021, Z-0973-2021, Z-1051-2021Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.Terminated

Other EOQ clearances

The newest 510(k)s for the same product code; their median review was 229 days; see every EOQ device

Record

Type
Traditional
Received
Jan 31, 2006
Review panel
Ear, Nose, Throat
Regulation
21 CFR 874.4680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files