Arca
Skip to content
Tools
Tools

Synthes (USA) Craniomaxillofacial Distraction System

External Mandibular Fixator And/Or Distractor. 510(k) K060138, from Synthes (Usa) (West Chester, PA). Product code MQN, class 2.

As of Nov 18, 2026, K060138 is an FDA 510(k) held by Synthes (Usa), substantially equivalent on Jun 2, 2006 after 134 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 134 days
  • Recalls: 2
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Synthes Craniomaxillofacial Distraction System (CMF Distraction System) is intended for use as a bone stabilizer and lengthening (and/or transport) device for correction of congenital deficiencies or post-traumatic defects of the mandibular body and ramus where gradual bone distraction is required. The Synthes CMF Distraction System is intended for single use only.

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K060138, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 16, 2014Z-2148-2014DePuy Synthes is initiating a recall of certain lots of the Craniomaxillofacial Distraction System (AB Distractor Bodies and BC Distractor Bodies) because they may reverse post-operatively.Terminated
Jun 12, 2013Z-1726-2013The firm initiated a voluntary recall of the BC Distractor Body which is part of the Craniomaxillofacial (CMF) Distractor System, due to a mis-alignment issue. There have been instances reported within the impacted lots…Terminated

Other MQN clearances

The newest 510(k)s for the same product code; see every MQN device

Record

Type
Traditional
Received
Jan 19, 2006
Review panel
Dental
Regulation
21 CFR 872.4760

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.