Arca
Skip to content

Protege Everflex Self-Expanding Biliary Stent System

Stents, Drains And Dilators For The Biliary Ducts. 510(k) K060057, from Ev3, Inc. (Plymouth, MN). Product code FGE, class 2.

As of Nov 17, 2026, K060057 is an FDA 510(k) held by Ev3, Inc., substantially equivalent on Mar 8, 2006 after 58 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Mar 8, 2006
  • Review time: 58 days
  • Recalls: 2
  • Cited as a predicate by: 4 devices

Cited as a Predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K060057, newest first; each opens FDA's recall record
Mar 31, 2010Z-1598-2010ev3 is conducting a voluntary recall on three lots of the Protg EverFlex" Biliary Stent System. They determined that a lot of 120mm Protg EverFlex" Biliary Stent system may contain a 150mm length Self-Expanding stent…Terminated
Nov 6, 2009Z-0478-2010ev3 has determined that a lot of 150mm Protg EverFlex Biliary Stent System contains a 100mm length Self-Expanding stent. The implantation of a stent 50mm shorter than expected may not cover the full length of the target…Terminated

Other FGE Clearances

The newest 510(k)s for the same product code; their median review was 162 days; see every FGE device

Record

Type
Special
Received
Jan 9, 2006
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5010

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files